SL-7603C Single Cable Vertical Flaming Tester

Ask Latest Price
Brand Name :SKYLINE
Model Number :SL-7603C
Place of Origin :CHINA
MOQ :1 UNIT
Price :SL-7603C Single Cable Vertical Flaming Tester
Payment Terms :TT , Western Union,paypal
Supply Ability :10 pcs/ month
Delivery Time :15 days after order confirmed
Packaging Details :standard exported package
Certification :SKYLINE
Specimen length :600±25mm
Gas flow range :0.1-1L/min
Area flow range :Area flow range
Burner :500W for standards power, 125±25mm for total height of the flame
Vertical burning chamber :W300xH1200xD450mm
Area of specimen :Total section area less than 0.5mm2
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description
SL-7603C Single Cable Vertical Flaming Tester

Standards

According to IEC60332-1-1:2004, IEC60332-1-2:2004, IEC60332-1-3:2004, EN50265-1, EN50265-2-1

Application

Applies to check thesingle insulation wire and cable vertical spreading test which total section area less than 0.5mm2

Technical parameters

Vertical burning chamber W300xH1200xD450mm
Gas Use high purity propane gas or liquefied petroleum gas (provide by user-self)
Specimen length 600±25mm
Area of specimen Total section area less than 0.5mm2
Testing temperature This device is located in a stuffiness room and the temperature keeping the 23±10℃ for testing
Burning time continuously setting between 0.1-999.9S, during the setting time flame could continuously burning the specimen
Control function Switch can choose to be manual or automatically according to the actual needed
Gas flow range 0.1-1L/min
Area flow range 0.1-1m3/h
Burner 500W for standards power, 125±25mm for total height of the flame
Temperature measurement system according to GB/T5169.14-2007 and IEC60695-11-03 standard requirements, including temperature copper (10g), K type thermocouple and diameter is 0.5 mm with stainless steel sheathed, stopwatch and thermometer etc.